Catalog Number SECX-10-7-40-135 |
Device Problems
Activation, Positioning or Separation Problem (2906); Material Deformation (2976)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/13/2016 |
Event Type
malfunction
|
Event Description
|
The physician was attempting to use a protege rx carotid self-expanding stent to treat a lesion in the ica.The device was inspected prior to use with no issues noted.No resistance noted during delivery of the device.During deployment of the stent in the ica, the delivery pusher shaft was fractured.Md retrieved the stent and used another product to complete procedure.No clinical sequelae reported.
|
|
Manufacturer Narrative
|
Evaluation summary : the protégé rx was received with approximately 2cm of the stent exposed/deployed.Outer was separated approximately 132cm from the distal tip.Due to the separated outer the stent could not be deployed.No damage was noted to the distal tip.Direct light was applied to the distal end of the catheter and red biological material was discovered within the outer lumen.(b)(4).
|
|
Search Alerts/Recalls
|