MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problems Unintended Arm Motion (1033); Unintended Movement (3026)
Patient Problem No Patient Involvement (2645)
Event Date 05/17/2016
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Return requested.Replacement articulating arm shipped to site 06/01/2016.No parts have been received by manufacturer for analysis.No further issues have been reported.
 
Event Description
A medtronic representative reported a site navigation system articulating arm that would not tighten properly.No further details regarding the damage, or how it occurred, were provided.There was no patient present when this issue was identified.
 
Manufacturer Narrative
The hardware investigation of the returned articulating arm found that the reported event was related to a mechanical issue.As reported, the arm will not hold when the handle has been tightened.The arm did not pass force testing.The reported event was confirmed.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
amos jarrette
826 coal creek circle
louisville, CO 80027-9710
7208902082
MDR Report Key5735526
MDR Text Key47743283
Report Number1723170-2016-01061
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeES
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 02/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/20/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberS7
Device Catalogue Number9733858
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received01/13/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/12/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-