MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE WATCHMAN® ACCESS SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE WATCHMAN® ACCESS SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL Back to Search Results
Model Number M635TS20060
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/25/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by mfr: the complaint device was not received for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
 
Event Description
It was reported a hemostasis valve leak occurred.A left atrial appendage (laa) closure procedure was being performed.A watchman access system was introduced inside the patient.After entering the laa, the dilator was removed and blood was leaking through the hemostasis valve which was unable to close.The physician continued the procedure applying thumb pressure to minimize the leak.Once the delivery system is inserted, the gap is closed and the leaking stops.The procedure was completed with this device.There were no patient complications reported and the patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN® ACCESS SYSTEM
Type of Device
SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5738795
MDR Text Key47844322
Report Number2134265-2016-05393
Device Sequence Number1
Product Code NGV
Combination Product (y/n)N
PMA/PMN Number
P130013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 05/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2019
Device Model NumberM635TS20060
Device Catalogue NumberTS-2006
Device Lot Number18796232
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/25/2016
Initial Date FDA Received06/21/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/28/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age75 YR
Patient Weight130
-
-