Analysis found foreign material inside the pump.The analysis lab was unable to achieve telemetry through the box via a model 8840 clinician programmer or lab view.No trauma to the box itself was observed by the analysis lab, and no alarms sounded.Upon opening the box and performing several tests in an attempt at telemetry, it was determined that the pump had a complete loss of power.X-ray analysis was performed and no anomalies that could cause a hard short and total battery depletion were observed.The pump was then opened and foreign material was observed on the (b)(4), now mobile.After opening the pump and removing the foreign material, (b)(4) was achieved.(b)(4) showed a hard short (por) had occurred and showed rising voltages in the minute history.The foreign material is extremely soft, colored with shades of tan, and contained white and blue fibers.The foreign material is consistent with a mixture of (b)(4).It is likely that the foreign material is biological in nature, and has been exposed to some sort of detergent, surfactant, or cosmetic.It is also possible that the foreign material is an (b)(4).
|