MAUDE Adverse Event Report: SYNTHES MEZZOVICO 4.5MM VA-LCP CURVED CONDYLAR PLATE/16 HOLE/336MM/RIGHT; IMPLANT,FIXATION DEVICE, CONDYLAR PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES MEZZOVICO 4.5MM VA-LCP CURVED CONDYLAR PLATE/16 HOLE/336MM/RIGHT; IMPLANT,FIXATION DEVICE, CONDYLAR PLATE Back to Search Results
Catalog Number 02.124.416
Device Problem Break (1069)
Patient Problems Loss of Range of Motion (2032); Non-union Bone Fracture (2369); No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
Device was used for treatment, not diagnosis.Patient initials are: (b)(6).Date of event: unknown.Additional product codes: hrs, hwc.Device is not expected to be returned for manufacturer review/investigation.(b)(4): a plate broke postoperatively after ambulation, in addition, the patient had a non-union and delayed healing.Revision surgery was required to remove the broken plate and intact hardware, and further treat the patient.Device history records was conducted.The report indicates that the lot # 8649709 manufacturing location: (b)(4), manufacturing date: 04.Oct.2013.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that a patient initially underwent for a plate fixation surgery at metaphyseal flare on (b)(6) 2016.The plate broke postoperatively after ambulation and it was also discovered that patient had a non-union and delayed healing.Patient underwent for revision surgery on (b)(6) 2016.The broken plate and unknown number of screws were explanted.Patient revised with retro grade femoral nail.The procedure was completed successfully with no surgical delay and any other medical intervention required.This complaint involves 1 device.Concomitant medical device reported: screws (part # unknown, lot # unknown, quantity # unknown).This report is 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM VA-LCP CURVED CONDYLAR PLATE/16 HOLE/336MM/RIGHT
Type of Device
IMPLANT,FIXATION DEVICE, CONDYLAR PLATE
Manufacturer (Section D)
SYNTHES MEZZOVICO
zona industriale 4
mezzovico CH680 5
SZ  CH6805
Manufacturer (Section G)
SYNTHES MEZZOVICO
zona industriale 4
mezzovico CH680 5
SZ   CH6805
Manufacturer Contact
terry callahan
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5740030
MDR Text Key47890802
Report Number1000562954-2016-10133
Device Sequence Number1
Product Code JDP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK110354
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 06/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number02.124.416
Device Lot Number8649709
Other Device ID Number(01)10886982042934(10)8649709
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/06/2016
Initial Date FDA Received06/21/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/04/2013
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age60 YR
Patient Weight118
-
-