• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES< PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER WITH AMC THROMBOSHIELD; SWAN-GANZ CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES< PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER WITH AMC THROMBOSHIELD; SWAN-GANZ CATHETER Back to Search Results
Model Number 774HF75J
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/07/2016
Event Type  malfunction  
Manufacturer Narrative
One catheter with attached monoject 1.5 cc limited volume syringe and two three-way stop cocks was returned for evaluation.A non-edwards contamination shield was located on the catheter body between 75 cm and 91 cm proximal from the catheter tip.No packaging or introducer was returned.The balloon was found to be torn from the proximal and distal bonding sites and the central area of balloon latex was missing.Bonding indication was observed from both the distal and proximal bonding sites.All through lumens were patent without any leakage or occlusion.No visible damage to the catheter body or returned syringe was observed.Balloon inflation testing was performed using returned syringe with 1.5 cc air.Visual examination was performed under microscope at 20x magnification and with the unaided eyes.A device history record review was completed and documented that device met all specifications upon distribution.Customer report of balloon inflation issue was confirmed.In addition, the balloon was found to be torn off between the bonds during evaluation.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.
 
Event Description
It was reported that the balloon did not inflate during use.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER WITH AMC THROMBOSHIELD
Type of Device
SWAN-GANZ CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES< PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
lynn selawski
1 edwards way
irvine, CA 92614
9497564386
MDR Report Key5740062
MDR Text Key47895996
Report Number2015691-2016-02006
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K040287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 04/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/21/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/16/2017
Device Model Number774HF75J
Device Lot Number60222214R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/19/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/07/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/05/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-