• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES USA GUIDEWIRE Ø1.25 W/THREAD-TIP W/TROCAR L1; WIRE,SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES USA GUIDEWIRE Ø1.25 W/THREAD-TIP W/TROCAR L1; WIRE,SURGICAL Back to Search Results
Catalog Number 292.620S
Device Problems Break (1069); Material Fragmentation (1261)
Patient Problems Failure of Implant (1924); No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
Event Date 06/05/2016
Event Type  Injury  
Manufacturer Narrative
Device was used for treatment, not diagnosis.(b)(4).Not implanted or explanted.Device is expected to be returned to synthes manufacturer for evaluation /investigation, but has yet to be received.(b)(6).Device history records was conducted.The report indicates that the: lot is either 9733073 or 9894439.The investigation could not be completed; no conclusion could be drawn, as no product was received.Lot 9733073 / unsterile 9682651.Manufacturing site: (b)(4), supplier: (b)(4), manufacturing date: 19.Nov.2015, expiry date: 01.Nov.2025.Part was first manufactured unsterile in balsthal under the lot 9682651, as the complaint is not related to sterilisation or packaging only the documents of the unsterile lot were reviewed: no ncrs were generated during production.Review of the device history record showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Lot 9894439 / unsterile 9881137, manufacturing site: (b)(4), supplier: (b)(4), manufacturing date: 05.April.2016, expiry date: 01.Mar.2026.Part was first manufactured unsterile in balsthal under the lot 9881137, as the complaint is not related to sterilisation or packaging only the documents of the unsterile lot were reviewed: no ncrs were generated during production.Review of the device history record showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes on an event in (b)(6) as follows: it was reported that a guide wire broke during surgery.The surgeon was previously unhappy about the drill bit and reported it as blunt, he was given another drill bit from another set, these are reused drill bits.He later reported that the guide wire was broken inside the patient and this was a direct result of using a blunt drill bit.This complaint involves 1 part.This report is 1 of 1 for (b)(4).
 
Manufacturer Narrative
(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Update 24 june 2016 - additional information received 15 june 2016: three articles were affected in total 2x guide wire and 1x drill bit.The broken off pieces were left in the patient.This report is 1 of 3 for (b)(4).
 
Manufacturer Narrative
Updated cia: two guide wires, 1.25mm, with threaded-tip with trocar, length 150mm (part number 292.620, lot 9894439 & 9733073) was reported as broken, but was not returned for further evaluation.Without product an investigation or a root cause cannot be performed/defined.The dhr review shows that the devices met the specifications at the time of manufacturing and distributing.Device used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Updated 02.August 2016:.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUIDEWIRE Ø1.25 W/THREAD-TIP W/TROCAR L1
Type of Device
WIRE,SURGICAL
Manufacturer (Section D)
SYNTHES USA
1302 wrights lane east
west chester PA 19380
Manufacturer Contact
terry callahan
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5740173
MDR Text Key47893951
Report Number2520274-2016-13242
Device Sequence Number1
Product Code LRN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PPREADM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 06/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/21/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number292.620S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received08/02/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-