• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS ENDOVASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Catalog Number FEM14120
Device Problem Positioning Failure (1158)
Patient Problems Vascular System (Circulation), Impaired (2572); No Known Impact Or Consequence To Patient (2692)
Event Date 05/27/2016
Event Type  Injury  
Manufacturer Narrative
The device history records are being reviewed.The event is currently under investigation.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant was unable to provide further patient or procedural details.
 
Event Description
It was reported that during stenting of a right upper arm fistula via upper arm access, the endovascular stent graft could not be deployed.The device was removed and three stent grafts of the same brand but smaller size were used instead to complete the procedure successfully.There was no reported patient injury.This is the same patient as reported in medwatch report # 9681442-2016-00173 and 9681442-2016-00174.
 
Manufacturer Narrative
The lot history records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met all specifications prior to shipment.There is no indication for a manufacturing-related failure.The condition of the returned device confirmed the reported deployment failure.The stent graft was found to be partially released and the outer sheath was found to be elongated indicating that increased friction must have affected on the delivery system during the attempt to deploy the stent graft.Potential contributing factors to the reported event have been considered.Previous investigations of similar complaints have been reviewed.This type of event may be associated with a difficult vessel anatomy leading to increased friction and subsequent partial stent graft deployment.Insufficient flushing of the device may be another contributing factor to the reported event.On the basis of the information available, a definite root cause for the reported failure could not be determined.The ifu states that the device must be flushed with sterile saline.Also the ifu states: "the safety and effectiveness of the device when placed across a tight bend (.) has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure." and "do not kink the delivery catheter or use excessive force during delivery to the target lesion." furthermore, the ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device." the reported application represents an off-label use of the device which is indicated for use in the treatment of in-stent restenosis in the venous outflow of hemodialysis patients dialyzing by either an arteriovenous (a/v) fistula or a/v graft.The safety and effectiveness of the device for a treatment as described has not been evaluated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Type of Device
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstrasse 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG -9681442
wachhausstrasse 6
karlsruhe 76227
GM   76227
Manufacturer Contact
daniela mueller
wachhausstrasse 6
karlsruhe 76227
GM   76227
0497219445
MDR Report Key5741634
MDR Text Key47936271
Report Number9681442-2016-00172
Device Sequence Number1
Product Code PFV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/21/2018
Device Catalogue NumberFEM14120
Device Lot NumberANZJ2778
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/27/2016
Initial Date FDA Received06/22/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/04/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/12/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age60 YR
-
-