• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX®ADULT DISPOSABLE ANESTHESIA BREATHING CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX®ADULT DISPOSABLE ANESTHESIA BREATHING CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) Back to Search Results
Catalog Number C45031503D-NL
Device Problem Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Smiths medical has received the sample device.A full evaluation is anticipated, but not yet begun as the device is currently in transit to the investigation site.Smiths medical will file a follow-up report detailing the results of the evaluation once it is completed.
 
Event Description
The customer stated that the portexadult disposable anesthesia breathing circuits breathing bag failed to hold air after 1 minute into use, no injury to the patient.No additional information was provided.
 
Manufacturer Narrative
One portex adult disposable anesthesia breathing circuits was received for investigation.The device was received opened and without its original packaging.A review of the device history record, of the reported lot, found no nonconformities or issues during manufacturing.Visual inspection was performed at a distance of 12" to 24" and under normal conditions of illumination.Inspection found no visual defects with the returned devices.During functional testing, the anesthesia breathing circuit and breathing bag were leak tested; no leaks were observed.An audit was conducted on 32 devices from a production lot.The devices were manufactured with the same process procedures as the reported device.The devices were visually inspected and leak tested.No defects were found during the audit of the 32 devices.Investigation was unable to confirm the reported issue and found that the device operated as intended.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX®ADULT DISPOSABLE ANESTHESIA BREATHING CIRCUITS
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
michele seliga
1265 grey fox rd
st. paul, MN 55112
7633833052
MDR Report Key5762466
MDR Text Key48573360
Report Number2183502-2016-01291
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 04/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/30/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/28/2019
Device Catalogue NumberC45031503D-NL
Device Lot Number3163366
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/10/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/11/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-