• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALERE SAN DIEGO, INC. INRATIO PT/INR TEST STRIPS; PROTHROMBIN TIME TEST

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALERE SAN DIEGO, INC. INRATIO PT/INR TEST STRIPS; PROTHROMBIN TIME TEST Back to Search Results
Model Number 100071
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem Test Result (2695)
Event Date 05/19/2016
Event Type  Injury  
Manufacturer Narrative
Investigation conclusion: investigation pending.
 
Event Description
Report received of discrepant inratio values.(b)(6).
 
Manufacturer Narrative
Investigation/conclusion it is indicated that the product is not returning for evaluation.Therefore, a review of the entire in-house testing history of the lot was performed.In-house testing on the reported strip lot met release criteria.The product performed as expected.A review of the manufacturing records for the lot did not uncover any non-conformances.The lot met release specifications.Although the customer was identified as having conditions that may impact the performance of the assay, a root cause could not be determined.Based on the information available, there is no indication of a product deficiency.No corrective action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INRATIO PT/INR TEST STRIPS
Type of Device
PROTHROMBIN TIME TEST
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
9975 summers ridge road
san diego CA 92121
Manufacturer Contact
ya-ling king
9975 summers ridge road
san diego, CA 92121
8588052084
MDR Report Key5763608
MDR Text Key48619754
Report Number2027969-2016-00482
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092987
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number100071
Device Lot Number373678A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/02/2016
Initial Date FDA Received06/30/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/11/2016
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-