Device was used for treatment, not diagnosis.Device is an instrument and is not implanted/explanted.A device history record review was performed for the subject device lot.The review showed that there were no issues during the manufacture of the product or any subcomponents that would contribute to this complaint condition.No non-conformances were generated during the production of the subject device.The subject device has been received and is currently in the evaluation process.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
Product investigation summary: one (1) t8 stardrive screwdriver shaft (part: 314.467 / lot: 9981812) was returned with a complaint stating that the tip was found in sterile processing twisted and over torqued.The complaint was able to be confirmed as the tip of the screwdriver was indeed twisted (approximately 30 degrees) in the direction of screw insertion.During the investigation, no product design issues or discrepancies were observed that may have contributed to the complaint condition.A visual inspection under 5x magnification and drawing review were performed as part of this investigation.No definitive root cause was able to be determined; however, the failure mode is consistent with the application of excessive torque.The design is adequate for its intended use when used and maintained as recommended.The t8 stardrive screwdriver shaft is a common trauma instrument noted in 30 system technique guides including: 2.4mm va lcp distal radius, compact distal radius, and modular clavicle plate.In each instance, the instrument is utilized for screw insertion/removal.Based on the complaint description, a specific system was unable to be identified.In the lcp distal radius system, the driver is available for use in final tightening when utilized in conjunction with a 0.8nm torque limiting attachment.The technique guide notes that use of the torque limiting attachment (tla) is mandatory when inserting locking screws into variable angle locking holes to ensure the adequate torque is applied.Final locking must be done manually using the tla.The relevant drawings for the returned instruments were reviewed (both from the time of manufacture and present revision): top-level and tip profile.The design, materials, and finishing processes were found to be appropriate for the intended use of the device.No definitive root cause was able to be determined; however the failure mode is consistent with the application of excessive torque.Device used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|