DEPUY SYNTHES POWER TOOLS BATT-HANDPIECE MOD F/TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 05.001.201
Device Problems Overheating of Device (1437); Loss of Power (1475); Defective Component (2292); Failure of Device to Self-Test (2937)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to device wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported by (b)(6) that during service and evaluation, it was observed that the bearing on the battery handpiece device was defective.It was further noted that the device failed pre-test for proper function of triggers, free movement and falling out protection (seel ring).It was noted in the service order that the device was powerless and overheating.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATT-HANDPIECE MOD F/TRS
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
terry callahan
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5775814
MDR Text Key49766500
Report Number8030965-2016-14153
Device Sequence Number1
Product Code GEY
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 06/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.201
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/13/2016
Initial Date FDA Received07/07/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/28/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-