Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Information (3190)
|
Event Date 06/20/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show the lot released with no recorded anomaly or deviation.There are warnings in the package insert that state that this type of event can occur: under possible adverse effects, number 10 states, "fatigue fracture of component can occur as a result of loss of fixation, strenuous activity, malalignment, trauma, non-union, or excessive weight." device remains in patient.
|
|
Event Description
|
Patient underwent a shoulder revision due to a fractured humeral tray.
|
|
Manufacturer Narrative
|
This follow-up report is being filed to relay corrected information.
|
|
Manufacturer Narrative
|
This follow-up report is being filled to relay additional information, which was unknown a the time of the initial medwatch.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Reported event was unable to be confirmed due to limited information received from the customer.Products were determined to not be compatible as competitor¿s products were used with zimmer biomet components.Record was reviewed and no discrepancies were found.Root cause is attributed to the device design.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Upon investigation and visual evaluation of the product it is confirmed that the stem has fractured off the humeral tray.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Search Alerts/Recalls
|