• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN ENTERAL FEEDING PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN ENTERAL FEEDING PUMP Back to Search Results
Model Number UNKNOWN ENTERAL FEEDING PUMP
Device Problems Device Alarm System (1012); Complete Blockage (1094); Precipitate in Device or Device Ingredient (1478); No Flow (2991)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/05/2016
Event Type  malfunction  
Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event Description
It was reported to covidien on 07/06/2016 that a customer had an issue with an enteral feeding pump.The customer states that the set had a pack of aminoleban with 200ml of hot water par 33ml/h.The set seemed to clogged and the pump was working (turning) but the formula did not flow forward.The tubing became cloudy and the formula seemed to be precipitated inside it.The pump did not alarm.At the 2nd feeding, empty alarm occurred but 180ml of water was left inside the bag.No patient harm.
 
Manufacturer Narrative
Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
A device history record review could not be performed because a lot number was not received with the complaint.Sample was received.The pump set could not fit with pump tester.A scar was issued to our supplier to address the root cause and action for solving this.A corrective action has been taken to address this issue.This complaint will be used for tracking and trending purposes.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERAL FEEDING PUMP
Type of Device
ENTERAL FEEDING PUMP
Manufacturer (Section D)
COVIDIEN
15 hampshire street
mansfield MA 02048
MDR Report Key5784843
MDR Text Key50098780
Report Number8040459-2016-00007
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUNKNOWN ENTERAL FEEDING PUMP
Device Catalogue NumberUNK EF
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/06/2016
Initial Date FDA Received07/12/2016
Supplement Dates Manufacturer Received07/06/2016
08/08/2018
Supplement Dates FDA Received08/01/2018
10/05/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-