ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) PROMOTE PLUS CRT-D; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3211-36
Device Problem Difficult to Interrogate (1331)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/07/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported prior to a normal device changeout due to elective replacement indicator, the device had difficulty completing interrogation.Multiple attempts even using the patient care system were not successful.The physician used a magnet alongside electrosurgical instruments when replacing the device.All functions were working well with the new device.The patient condition was well before and after the procedure.
 
Manufacturer Narrative
The reported inability to interrogate was not confirmed in the laboratory.The device was tested on the bench and no anomaly was found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMOTE PLUS CRT-D
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jaime chavez
645 almanor avenue
sunnyvale, CA 94085
8184934022
MDR Report Key5788034
MDR Text Key49393715
Report Number2938836-2016-05983
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2012
Device Model NumberCD3211-36
Device Lot Number3383188
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/07/2016
Initial Date FDA Received07/12/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/18/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/06/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-