Model Number P143
Device Problems Device Alarm System (1012); Failure to Capture (1081); High impedance (1291); Pocket Stimulation (1463); Device Sensing Problem (2917)
Patient Problems Muscle Stimulation (1412); No Consequences Or Impact To Patient (2199)
Event Date 04/05/2016
Event Type  malfunction  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this patient stated the device was beeping sixteen times once a day.Boston scientific technical services (ts) reviewed the available data and indicated there were alerts for the right atrial (ra) lead for out of range amplitude measurements and left ventricular (lv) lead for out of range impedance measurements.The patient was instructed to contact their clinic for review.The patient had a history of loss of capture and diaphragm stimulation in other polarities, therefore, the programming could not be changed.The lv impedance measurement trigger was increased and the patient will continue to be followed appropriately.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENERGEN
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key5796176
MDR Text Key50347382
Report Number2124215-2016-10358
Device Sequence Number1
Product Code LWS
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial
Report Date 06/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/18/2013
Device Model NumberP143
Other Device ID NumberENERGEN CRT-D
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 06/21/2016
Initial Date FDA Received07/14/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/21/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age74 YR
-
-