ATRICURE, INC. COBRA FUSION 150 ABLATION MINIMALLY INVASIVE SYSTEM Back to Search Results
Model Number 001-700-001MI
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/08/2015
Event Type  malfunction  
Manufacturer Narrative
Complaint number: (b)(4).The device was received and sent to atricure engineering for analysis.The complaint was confirmed, the magnet cap had failed at the front face resulting in the magnet pulling from the distal end of the probe.
 
Event Description
It was reported that the magnet in the end of the fusion 150 device became dislodged and attached itself to the end of the magnetic introducer.The case was completed with the same device, patients outcome was not affected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBRA FUSION 150 ABLATION MINIMALLY INVASIVE SYSTEM
Type of Device
COBRA FUSION 150 ABLATION MINIMALLY INVASIVE SYSTEM
Manufacturer (Section D)
ATRICURE, INC.
6217 centre park drive
west chester OH 45069 3886
Manufacturer Contact
ranjana iyer
6217 centre park drive
west chester, OH 45069-3886
5137555320
MDR Report Key5799625
MDR Text Key49754642
Report Number3003502395-2015-00052
Device Sequence Number1
Product Code OCL
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K113475
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 12/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/01/2018
Device Model Number001-700-001MI
Device Catalogue Number001-700-001MI
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received12/09/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/20/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-