• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; N/A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; N/A Back to Search Results
Catalog Number A1059
Device Problems Out-Of-Box Failure (2311); Device Operates Differently Than Expected (2913); Material Integrity Problem (2978)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Event Description
Presents a flaw in the cushion that covers the pin in the torque screw.There was no patient prepped for surgery and no revision/medical intervention required.Out of box failure reported.
 
Manufacturer Narrative
Integra has completed their internal investigation on 8/23/2016.The investigation included: methods: evaluation of actual device.Review of device history records.Results: evaluation of device: the skull clamp was received with the helicoil outside of the ratchet assembly.Device history record reviewed for this product id lot code/work order: 159/135911 manufactured on 12/03/15 show no abnormalities related to the reported failure.This device passed all required inspection points with no associated mrr¿s, variances or rework.No service history is on file for this device.In summary: the helicoil displacement issue reported on the returned devices was able to be confirmed by integra engineering and quality.Capa has been opened to investigate this failure.The skull clamp¿s ratchet extension has a detailed assembly instruction that outlines the process of installing the helicoil.The last step in the work instructions requires the operator to use a ¾-16 thread gage to verify fit.All products in question were subjected to these quality controls during manufacture.These will be monitored for trending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
N/A
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH 45227
Manufacturer Contact
rowena bunuan
311 enterprise drive
plainsboro, NJ 08536
6099362393
MDR Report Key5804200
MDR Text Key50731046
Report Number3004608878-2016-00191
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/19/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1059
Device Lot Number159
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received08/23/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/03/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-