Integra has completed their internal investigation on 8/23/2016.The investigation included: methods: evaluation of actual device.Review of device history records.Results: evaluation of device: the skull clamp was received with the helicoil outside of the ratchet assembly.Device history record reviewed for this product id lot code/work order: 159/135911 manufactured on 12/03/15 show no abnormalities related to the reported failure.This device passed all required inspection points with no associated mrr¿s, variances or rework.No service history is on file for this device.In summary: the helicoil displacement issue reported on the returned devices was able to be confirmed by integra engineering and quality.Capa has been opened to investigate this failure.The skull clamp¿s ratchet extension has a detailed assembly instruction that outlines the process of installing the helicoil.The last step in the work instructions requires the operator to use a ¾-16 thread gage to verify fit.All products in question were subjected to these quality controls during manufacture.These will be monitored for trending.
|