MAKO SURGICAL CORP. MCK FEMORAL-LM-RL-SZ 3; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Catalog Number 180503
Device Problem Failure To Adhere Or Bond (1031)
Patient Problem Pain (1994)
Event Date 06/22/2016
Event Type  Injury  
Manufacturer Narrative
An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Additional information has been requested.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event Description
Original date of surgery was (b)(6) 2015.Patient presented with pain during a follow up appointment.Doctor said patient was revised to a tka for possible aseptic loosening and no radiolucent lines on x-ray.
 
Manufacturer Narrative
An event regarding loosening involving an mako femoral component was reported.The event was not confirmed.Method & results: device evaluation and results: a visual, functional and dimensional inspection could not be performed as the device was not returned.Medical records received and evaluation: no medical records or x-rays were made available for evaluation.Device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.Complaint history review: there has been no other event for the lot referenced.Conclusions: the event could not be confirmed nor the root cause determined because the devices were not returned for evaluation and insufficient medical information was provided.If the devices and/or additional information are received, this investigation will be reopened and re-evaluated.
 
Event Description
Original date of surgery was (b)(6) 2015.Patient presented with pain during a follow up appointment.Doctor said patient was revised to a tka for possible aseptic loosening and no radiolucent lines on x-ray.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCK FEMORAL-LM-RL-SZ 3
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5808193
MDR Text Key49963873
Report Number3005985723-2016-00225
Device Sequence Number1
Product Code NPJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/30/2020
Device Catalogue Number180503
Device Lot Number237299K-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/23/2016
Initial Date FDA Received07/20/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received09/26/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/17/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age68 YR
-
-