DEPUY SYNTHES POWER TOOLS SMALL BATTERY DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.010 |
Device Problems
Break (1069); Sticking (1597); Naturally Worn (2988)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/09/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was found that the upper trigger was sticky, and the trigger components were worn.It was also noted that the electric control unit¿s (ecu) wire insulation was damaged.The assignable root cause was determined to be due to wear from normal use and servicing over time.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported that the small battery drive device would not grad attachments.During in-house engineering evaluation it was found that the upper trigger was sticky, and the trigger components were worn.It was also noted that the units electric control unit's ecu) wire insulation was damaged.It is unknown whether the event occurred during surgery.It was not reported if there was a delay in the procedure of if a spare device was available for use.There were no patient or user injuries reported.It was reported there was no medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|