• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - IRVINE FOLFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - IRVINE FOLFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number D2C4009K
Device Problem Cut In Material (2454)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that there was a cut in the tubing near the control module of a large volume foflusor.This was observed during filling.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
(b)(4).Device manufacture date: february 19, 2016 ¿ february 20, 2016.The device was received for evaluation.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.A visual inspection was performed and revealed a surface cut located on a segment of the tubing.The surface cut was measured to be approximately 2.11 mm in length.There was no leak observed despite the surface cut on the tubing.The reported condition was verified.The cause of the condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue, bldg 3
irvine CA 92614
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue, bldg 3
irvine CA 92614
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5813043
MDR Text Key50391138
Report Number1416980-2016-12129
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date02/01/2018
Device Catalogue NumberD2C4009K
Device Lot Number16B034
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/29/2016
Initial Date FDA Received07/21/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/19/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-