• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS STERNAL SAW SYSTEMS; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL-SARNS STERNAL SAW SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS STERNAL SAW SYSTEMS; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL-SARNS STERNAL SAW SYSTEMS Back to Search Results
Model Number 15665
Device Problem Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/30/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Evaluation is in progress, but not yet concluded.This complaint is related to emdr #1828100-2016-00513.The user facility¿s biomedical engineer (biomed) indicated that the flexible drive cable did connect for him but was very difficult to connect.Sterilization group said that in the past it was not difficult to connect the saw and cable.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the flexible drive cable would not connect to the sternal saw and there was a bent blade guard/protector (refer to emdr #1828100-2016-00513).As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was confirmed per the user facility biomedical engineer.The sternal saw (refer to emdr #1828100-2016-00513) and cable were returned for repair.The customer decided not to repair the unit.The unit was returned to the customer not repaired and marked not for clinical use.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS STERNAL SAW SYSTEMS
Type of Device
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL-SARNS STERNAL SAW SYSTEMS
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key5815576
MDR Text Key51013958
Report Number1828100-2016-00514
Device Sequence Number1
Product Code GFA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K935391
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 10/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number15665
Device Catalogue Number15665
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/28/2016
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-