Model Number G141
Device Problems High impedance (1291); Over-Sensing (1438); Ambient Noise Problem (2877)
Patient Problem Patient Problem/Medical Problem (2688)
Event Date 03/28/2016
Event Type  malfunction  
Manufacturer Narrative
No additional information is available.If additional information becomes available the report will be updated at that time.
 
Event Description
Boston scientific received information that there was oversensing of noise on the right atrial (ra) channel.Upon review by boston scientific technical services (ts) they noticed there was oversensing of the minute ventilation signal which can indicate an issue with the ra lead or device.Further review found one high out of range pacing impedance measurement for the other manufacturer's ra lead and this device.Several spikes in the ra impedance measurement were also seen.Ts suggested further evaluation.The field representative noted that the patient was seen in the clinic by an electrophysiologist (ep) physician and at this time they will not be taking any further action with the device.It was also noted that the patient has mental health issues that make her unreliable for follow up appointments.The ra lead and device remain in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN
Type of Device
IMPLANTABLE CHF GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key5816209
MDR Text Key50297362
Report Number2124215-2016-11333
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534553
UDI-Public(01)00802526534553(17)20171006
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 05/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/06/2017
Device Model NumberG141
Other Device ID NumberINOGEN CRT-D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received05/17/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/06/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age57 YR
-
-