Model Number 3721 |
Device Problem
Temperature Problem (3022)
|
Patient Problem
Burning Sensation (2146)
|
Event Date 06/29/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).This charger model was associated a field correction.The investigation for (b)(4) associated with heating while charging (pocket heating) concluded that the charger was capable of transferring energy to the ipg at a rate that would cause heating of the ipg and/or charging wand of sufficient elevated temperature to cause pain and burns.The heating while charging was determined to be exacerbated by off-axis charging of shallow implanted ipgs and that all chargers were capable of elevated heating.(b)(4).
|
|
Event Description
|
Device 2 of 2.Reference mfr.Report: 1627487-2016-03679.It was reported the patient experienced heating at the ipg site while charging.Surgical intervention is planned to address the issue.On (b)(6) 2012 st.Jude medical, neuromodulation division, sent field action letters to patients related to heating while charging and raised awareness of this issue to patients.An increase in prior non-reported heating while charging events and other non-reported events was expected.
|
|
Manufacturer Narrative
|
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Event Description
|
Device 2 of 2: reference mfr.Report: 1627487-2016-03679.Follow-up identified surgical intervention was undertaken on (b)(6) 2016 to explant and replace the ipg.The issue has resolved.
|
|