Model Number 420591
Device Problem Fluid/Blood Leak (1250)
Patient Problem Bacterial Infection (1735)
Event Type  Injury  
Manufacturer Narrative
Based on the available information, this event is deemed to be a serious injury.A previous investigation covers all the issues of pinholes, partial weld, underweld, overweld on collar weld and pouch film separate from mass.The product was manufactured under device specification.The crew requirements and responsibilities, process parameters, quality and in-process inspections, line operations, process troubleshooting and relevant documents to the process were run according to the process instructions.The process requirements results were documented in the product batch records and the product was packaged and labeled under the packaging and labeling specification.A non-conformance was generated to investigate the issue of pinholes, partial weld, underweld, overweld on collar weld and pouch film separating the from mass.No additional investigation is required.Additional information has been requested for patient/event details.Should additional information become available, a follow-up report will be submitted.Note: there is another case associated with this complaint.A separate 3500a form has been completed for the other case.(b)(4).
 
Event Description
It was reported by a hospital worker, that the pouch leaked while worn by a patient, resulting in a bacterial infection.It was also stated that this prolonged the recovery time for the patient.No further details have been provided.
 
Manufacturer Narrative
A batch record review indicates no discrepancies.A photograph has been received for this complaint and has been evaluated.Evaluation of the photograph reveals the product is a convatec product and is defective.This deficiency is covered under the nonconformance (nc) reported on the previous submission, mfr report #9618003-2016-00034, submitted to the fda on july 27, 2016.No additional patient/event details have been provided to date.Should additional information become available a follow-up report will be submitted.Reported to the fda on october 11, 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVELIFE ONE-PIECE DRAINABLE POUCH
Type of Device
POUCH, COLOSTOMY
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC, INC.
carretera sanchez, km 18.5
parque industrial itabo, s.a.
haina, san cristobal 33102
DR  33102
Manufacturer Contact
jeanette johnson
7900 triad center drive
suite 400
greensboro, NC 27409
3365424681
MDR Report Key5825644
MDR Text Key50531710
Report Number9618003-2016-00035
Device Sequence Number1
Product Code EZQ
Combination Product (y/n)N
Reporter Country CodeMY
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 07/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/02/2020
Device Model Number420591
Device Lot Number5B00547
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/01/2016
Initial Date FDA Received07/27/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/11/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/05/2015
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-