• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS, INC MERIT CUSTOM KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS, INC MERIT CUSTOM KIT Back to Search Results
Catalog Number K05-YSI032B
Device Problem Device Packaging Compromised (2916)
Patient Problem No Patient Involvement (2645)
Event Date 06/21/2016
Event Type  malfunction  
Manufacturer Narrative
One device was returned for evaluation.The investigation is in progress.A follow up will be submitted when the evaluation is complete.
 
Event Description
The distributor reported the sealing had run over the edge of a pouch inside kit.This was identified during incoming inspection and was not sent to a user facility.
 
Manufacturer Narrative
One device was returned for evaluation.The device was visually inspected and ingress testing was performed.A breach was found in the sterile barrier.The complaint is confirmed.The root cause is attributed to the manufacturing process.The device history record was reviewed and no exception documents were found.The complaint database was reviewed and no similar complaints for this lot number were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERIT CUSTOM KIT
Type of Device
CUSTOM KIT
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS, INC
1600 west merit parkway
south jordan UT 84095
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS, INC
1600 west merit parkway
south jordan UT 84095
Manufacturer Contact
jerry mcphie
1600 west merit parkway
south jordan, UT 84095
8012084491
MDR Report Key5829958
MDR Text Key50657822
Report Number1721504-2016-00127
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K884913
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2019
Device Catalogue NumberK05-YSI032B
Device Lot NumberH967150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/04/2016
Initial Date FDA Received07/28/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/03/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/14/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-