• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INVACARE FLORIDA OPERATIONS CS BED 9153650455; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVACARE FLORIDA OPERATIONS CS BED 9153650455; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Model Number G5510
Device Problems Device Inoperable (1663); Sparking (2595)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Should additional information become available a supplemental record will be filed.
 
Event Description
Sparks were seen coming from the foot section of the bed and then bed had no function at all.
 
Manufacturer Narrative
Additional/updated information was added to reflect the device being returned to the manufacturer for evaluation, however subsequent testing could not verify the complaint, and the underlying cause of the complaint issue could not be determined.Per the expanded evaluation report, no sparks were observed coming from any of the electronic components on the foot section during bed operation.Both the head and foot actuators operated properly when tested.However, the hi-lo actuator could not be operated using the returned hand pendant due to a broken/missing connector pin.
 
Event Description
Sparks were seen coming from the foot section of the bed and then bed had no function at all.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS BED 9153650455
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer (Section G)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer Contact
kevin guyton
one invacare way
elyria, OH 44035
8003336900
MDR Report Key5833480
MDR Text Key51162807
Report Number1031452-2016-03139
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Medical Equipment Company Technician/Representative
Remedial Action Replace
Type of Report Initial,Followup
Report Date 08/31/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/29/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberG5510
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received08/31/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-