• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-723NAS
Device Problems Failure to Prime (1492); Device Displays Incorrect Message (2591); Malposition of Device (2616)
Patient Problem Hyperglycemia (1905)
Event Date 06/03/2014
Event Type  malfunction  
Manufacturer Narrative
Analysis found the unit alarmed compromised force sensor system due to a loose/protruded drive support disk.Note: this is a remediation mdr.Medtronic diabetes implemented revised mdr reportability criteria effective on july 1, 2014.Subsequently, medtronic diabetes conducted a one year retrospective review of complaints.This event was retrospectively identified to be reportable based on the revised mdr reportability criteria.
 
Event Description
The customer reported via phone call that the insulin pump received a compromised force sensor system alarm.His blood glucose was 370 mg/dl at the time of incident.The customer stated that the insulin pump was stuck in prime rewind.He stated the insulin pump was not dropped or bumped.He stated that the drive support cap was recessed.The customer was advised that the insulin pump will need to be replaced.The customer was advised to disconnect from the insulin pump and revert to back-up plan.The insulin pump will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key5833959
MDR Text Key51736948
Report Number3004209178-2016-61563
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 06/04/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-723NAS
Device Catalogue NumberMMT-723NAS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/04/2014
Initial Date FDA Received07/29/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age70 YR
Patient Weight91
-
-