• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIAGNOSTICS DIVISION CELL-DYN EMERALD CLEANER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIAGNOSTICS DIVISION CELL-DYN EMERALD CLEANER Back to Search Results
Catalog Number 09H46-02
Device Problems Nonstandard Device (1420); Low Test Results (2458)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A product recall letter was issued to all cell-dyn emerald customers who have received cell-dyn emerald cleaner (lots 6853, 6901, 6953, 6991, 7024, 7027, 7044, 7082, 7110 and/or 7119).The letter informs the customer of the issue regarding the potential for out of range controls.The letter instructs the customer to discontinue use of the suspected lot and destroy any remaining inventory.The customer is instructed that if they do not have replacement material available, they can continue to use the lot until they receive replacement material as long as their controls are in range.The customer is instructed to perform the decontamination procedure per the operators manual.The cause for the qc material out-of-range low issue has been traced to the manufacturing process for raw material used in the cell-dyn emerald cleaner.
 
Event Description
The customer was experiencing rbc and plt control values out of range low with the use of cell-dyn emerald cleaner reagent lot number 7082.No adverse impact to patient management was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL-DYN EMERALD CLEANER
Type of Device
CELL-DYN EMERALD
Manufacturer (Section D)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
Manufacturer (Section G)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key5834551
MDR Text Key51767374
Report Number2919069-2016-01829
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081495
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Recall
Type of Report Initial
Report Date 07/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2018
Device Catalogue Number09H46-02
Device Lot Number7082
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/13/2016
Initial Date FDA Received07/29/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/14/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction Number2919069-03/24/16-001-R
Patient Sequence Number1
-
-