U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC CD HORIZON SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC CD HORIZON SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Catalog Number 7578302
Device Problem Difficult to Remove (1528)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/07/2016
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: 7578302, lot no: em15l033, qty: 8; product id: 7578302, lot no: em15h050, qty: 4.The product is not returned to manufacturer for evaluation therefore cause of event cannot be determined.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that peri operative, extender was not removable from the screws.It was not possible to pull the extender into the eject position.No patient's complications were reported.
 
Manufacturer Narrative
Product analysis: all 12 extenders were evaluated and all of them have same results which is as follows: functional bench evaluation of the instrument did not identify significant issues with opening, closing, or releasing a sample mas head from the instrument retention features.During in vivo usage, surrounding biological material or technique could potentially exert external forces on the instrument, and increase the force required to remove the instrument from the mas head.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
greg anglin
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key5836071
MDR Text Key50814236
Report Number1030489-2016-02227
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K122862
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 08/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number7578302
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/07/2016
Initial Date FDA Received08/01/2016
Supplement Dates Manufacturer ReceivedNot provided
08/03/2016
Supplement Dates FDA Received08/08/2016
09/23/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/07/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age79 YR
-
-