On (b)(6) 2016, there was an incident in the operating room involving the a1059 mayfield modified skull clamp attachments with respect to the open jackson beds.A (b)(6) male patient was in prone position undergoing a cervical laminectomy procedure.His head was secured to the mayfield skull clamp.As the surgical resident proceeded to attach the skull clamp to the open table adaptor, the patients head slipped from the mayfield attachment and his face hit the mayfield skull clamp.The incident caused a small superficial laceration and delayed the surgery by 20 minutes.The laceration was treated with topical bacitracin ointment and followed up by the surgical team.The patient had a good outcome.Adult disposable skull pins were used.The lot number and catalog number for these specific pins was not available.
|
Integra has completed their internal investigation on 26 aug 2016.The investigation included: methods: evaluation of actual device.Review of device history records.Review of complaint history.Results: evaluation of device: the complaint was not confirmed and no issues were observed.With respect to the returned unit it has passed all specific functional testing requirements, except for the lock having rotational movement, this would not have caused a slippage; when unit is properly positioned and put under pressure unit would not have slipped.General maintenance and cleaning required.This device was manufactured on 08/23/2014 and a review of the dhr containing lot code 147 and serial number (b)(4) showed that this lot passed the required inspection points without mrrs or variances.There were 2 reworks.No manufacturing or design related trend has been identified.Conclusion: in summary we are unable to confirm or duplicate the end users experience.The returned unit passed all functional testing requirements, however general maintenance is required.
|