MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) PROMOTE ACCEL ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) PROMOTE ACCEL ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3215-36
Device Problems Premature Discharge of Battery (1057); Failure to Convert Rhythm (1540); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problems Fainting (1847); No Known Impact Or Consequence To Patient (2692); Shock from Patient Lead(s) (3162)
Event Date 07/14/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that a patient presented in the hospital after receiving inappropriate shock.During the arrhythmia episode atp failed to stop the vt, which accelerated into the vf zone and shock was delivered.The episode was long and the patient fainted.The patient was recovering in the hospital and was stable.The physician decided not to make any programming changes, instead, patient's medication will be adjusted.
 
Event Description
New info received: the device was explanted and replaced due to battery depletion.Procedure was completed without any complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMOTE ACCEL ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jaime chavez
645 almanor avenue
sunnyvale, CA 94085
8184934022
MDR Report Key5842091
MDR Text Key50998678
Report Number2938836-2016-06808
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
PP030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2011
Device Model NumberCD3215-36
Device Lot Number3109919
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/14/2016
Initial Date FDA Received08/02/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/08/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/09/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-