• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NOVA BIOMEDICAL NOVA WAS GLUCOSE TEST STRIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NOVA BIOMEDICAL NOVA WAS GLUCOSE TEST STRIPS Back to Search Results
Lot Number 1020416003
Device Problems Defective Component (2292); High Test Results (2457)
Patient Problem Hyperglycemia (1905)
Event Date 08/04/2016
Event Type  Injury  
Event Description
Pt had appt with pcp this morning and c/o of getting very high bs results on her meter and asked for her meter to be checked.Pt reports that for 1 week her meter has been registering bs over 400, when checking pt's test strips with clinic meter using control solution, results were out of range, results were 404 mg/dl, control ranges should be 82-127mg/dl, found test strips defective, checked meter and working properly.Pt was given prescription for new test strips and instructed to go to the pharmacy for new test strips, also reviewed with pt safe storage of glucometer/test strips with handouts given, pt verbalized understanding.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVA WAS GLUCOSE TEST STRIPS
Type of Device
NOVA WAS GLUCOSE TEST STRIPS
Manufacturer (Section D)
NOVA BIOMEDICAL
walthan MA 02454
MDR Report Key5856061
MDR Text Key51741536
Report NumberMW5063909
Device Sequence Number1
Product Code CGA
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/31/2018
Device Lot Number1020416003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/04/2016
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age69 YR
Patient Weight66
-
-