• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HOL ML 5MM ENDO APPLIER; APPLIER,SURGICAL, CLIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HOL ML 5MM ENDO APPLIER; APPLIER,SURGICAL, CLIP Back to Search Results
Catalog Number 544965
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/27/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device investigation has not been submitted at this time.The manufacturer will continue to monitor and trend related events.
 
Event Description
The clips fell from applier easily and some clips were dropped in the patient during the procedure.All the clips were retrieved.The patient's condition was reported as fine.
 
Manufacturer Narrative
(b)(4).The dhr for the instrument in question, was reviewed and found completely without any irregularities.This instrument was manufactured (b)(4) in may of 2015.The returned instrument was evaluated and found that the jaws are aligned with each other and the tip set as received is correct per print specs.Further evaluation showed that this instrument picks up, retains, closes and releases clips both with and without the use of silastic test tubing as required of its function.We are unable to validate the alleged complaint since we are unable to duplicate the alleged issue.Parts were 100% visually inspected and tested at the (b)(4) facility before instruments were sent to customer.No irregularities were found and or reported at the time of inspection and assembly of the product as this is a standardized process for all instruments manufactured at this facility.No corrective action required at this time.
 
Event Description
The clips fell from applier easily and some clips were dropped in the patient during the procedure.All the clips were retrieved.The patient's condition was reported as fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOL ML 5MM ENDO APPLIER
Type of Device
APPLIER,SURGICAL, CLIP
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key5876328
MDR Text Key52107344
Report Number3011137372-2016-00225
Device Sequence Number1
Product Code GDO
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number544965
Device Lot Number06C1400239
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/04/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/19/2016
Initial Date FDA Received08/15/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/19/2016
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-