• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VOLCANO CORPORATION VOLCANO S5/S5I INTRAVASCULAR ULTRASOUND IMAGING AND PRESSURE SYSTEM; SYSTEM,IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VOLCANO CORPORATION VOLCANO S5/S5I INTRAVASCULAR ULTRASOUND IMAGING AND PRESSURE SYSTEM; SYSTEM,IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number COREMB120
Device Problem Inaccurate Synchronization (1609)
Patient Problem No Information (3190)
Event Date 05/20/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This case was reviewed and investigated according to philips volcano policy.Attempts to obtain patient information have been unsuccessful.Attempts to obtain the patient information were made via email and phone.All reasonably known patient information is included in this report.The system was analyzed and repaired at the facility; it was not returned to the manufacturer.The system was refurbished 02/19/2016.The cause of the pd pa waveform time shift issue is under investigation in accordance with the manufacturer's policy.A review of complaint cases submitted for this asset was performed and no other similar complaints were reported on this asset for this failure.This complaint will be monitored as part of complaint data analysis.
 
Event Description
The manufacturer's coronary account manager (cam) liaison for the facility reported the facility was experiencing intermittent ffr issues.The manufacturer's field service engineer (fse) visited the site and confirmed the issue by recalling previous cases and seeing a shift in the pd heart cycle, a waveform delay.The fse repaired the system by reloading software.A service checklist was performed to test and inspect the system.The fse noted the system was operating as intended.
 
Manufacturer Narrative
(b)(4).This case was reviewed and investigated according to the manufacturer's policy.Based on the previous assessment of this pa-pd shift complaint, the initial information available suggested that if pa-pd waveform shift occurs, inaccurate measurements could contribute to incorrect treatment decisions.The manufacturer's operator's manual requires use only by trained medical personnel and cautions that "federal law restricts this device to sale by or on the order of a physician (or properly licensed practitioner)." additionally, the operator's manual states that the ffr option on the manufacturer's system should be used only by physicians who are thoroughly trained in cardiac or peripheral vessel catheterization procedures.The fractional flow reserve (ffr) option on the manufacturer's system requires two pressure signals (pd and pa) to execute normalization and calculate results.The manufacturer's pressure wire, provides the distal pressure signal (pd).The physiologic aortic pressure signal (pa) comes from a sensor that is part of the hemodynamic monitoring system, a non-manufacturer's product.The aortic pressure sensor and attached guide catheter are set-up and positioned by the physician separate from manufacturer's pressure wire.Normalization is performed with the wire's pressure sensor located at the tip of the guide catheter within the coronary artery.Therefore, the physician initiates normalization once they have confirmed the placement of the pressure wire.Based on the above described factors (pd coming from volcano's pressure wire, pa coming from a non-manufacturer's product, and physician's initiation of normalization upon confirmation of placement of pressure wire), the physiologic aortic pressure signal can contain content that may cause a one cardiac cycle shift.Refer to manufacturer's operator's manual, section acquire ffr pressure measurements for warning on discontinuing use of the system, and contacting manufacturer's technical support if normalization fails to establish a pd/pa ratio of 1.0 after multiple attempts.Manufacturer's product manages such a shift by displaying the normalized waveforms so that the user can detect a one-cardiac cycle shift.Manufacturer's software (ffr v2.4 software) allows the user to evaluate the results of the normalization process and make adjustments if needed.The one-cardiac cycle shift is not caused by manufacturer's software and this not a product performance issue, as the product is working as designed the appropriate action is to review the presented waveforms and re-normalize with the goal of achieving a shift free normalization.Manufacturer's ffr v2.4 operator's manual has instructions in that if the user is unable to establish a post-normalization pd/pa ratio of 1.0 with overlapping waveforms after several attempts, they are instructed to contact manufacturer's technical support.As described in this document, ffr version v2.4 is performing as intended.
 
Event Description
This event is being submitted to correct and clarify information provided in the initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOLCANO S5/S5I INTRAVASCULAR ULTRASOUND IMAGING AND PRESSURE SYSTEM
Type of Device
SYSTEM,IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
VOLCANO CORPORATION
2870 kilgore road
rancho cordova CA 95670
Manufacturer (Section G)
VOLCANO CORPORATION
2870 kilgore road
rancho cordova CA 95670
Manufacturer Contact
tom brennan
3721 valley centre dr #500
san diego, CA 92130
8587641320
MDR Report Key5887658
MDR Text Key52500802
Report Number2939520-2016-00058
Device Sequence Number1
Product Code IYO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123898
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Remedial Action Repair
Type of Report Initial,Followup
Report Date 05/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCOREMB120
Device Catalogue Number400-0100.01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/19/2016
Initial Date FDA Received08/18/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/31/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/19/2016
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-