• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM ONBOARD BATTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM ONBOARD BATTERY Back to Search Results
Catalog Number 295025-001
Device Problems Failure to Charge (1085); Charging Problem (2892); Device Operates Differently Than Expected (2913)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 08/08/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The customer reported that the patient reported that the freedom onboard battery rapidly loses its charge.The customer also reported that the patient's onboard battery was exchanged.There was no reported patient impact.This alleged failure mode poses a low risk to the patient because although the freedom onboard battery has a loses its charge, it did not prevent the freedom driver from performing its life-sustaining functions.In addition, patients are provided with several onboard batteries, and the freedom driver has a redundant power source of external wall power.The freedom onboard battery will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.
 
Manufacturer Narrative
The freedom onboard battery was returned to syncardia for evaluation.The customer-reported issue was confirmed during testing; the onboard battery could not charge or provide power output after removal from a freedom onboard battery charger.A review of the system management bus (smbus) data revealed a fault code that indicated that the onboard battery was permanently disabled due to a pack over-voltage condition, causing its safety monitor firmware to permanently disable its output.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4).
 
Event Description
The customer reported that the patient reported that the freedom onboard battery rapidly loses its charge.The customer also reported that the patient's onboard battery was exchanged.There was no reported patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM ONBOARD BATTERY
Type of Device
BATTERY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
carole marcot
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key5888749
MDR Text Key53589100
Report Number3003761017-2016-00274
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date 08/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number295025-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/08/2016
Initial Date FDA Received08/19/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/16/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age65 YR
-
-