Catalog Number 295025-001 |
Device Problems
Failure to Charge (1085); Charging Problem (2892); Device Operates Differently Than Expected (2913)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/08/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
The customer reported that the patient reported that the freedom onboard battery rapidly loses its charge.The customer also reported that the patient's onboard battery was exchanged.There was no reported patient impact.This alleged failure mode poses a low risk to the patient because although the freedom onboard battery has a loses its charge, it did not prevent the freedom driver from performing its life-sustaining functions.In addition, patients are provided with several onboard batteries, and the freedom driver has a redundant power source of external wall power.The freedom onboard battery will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.
|
|
Manufacturer Narrative
|
The freedom onboard battery was returned to syncardia for evaluation.The customer-reported issue was confirmed during testing; the onboard battery could not charge or provide power output after removal from a freedom onboard battery charger.A review of the system management bus (smbus) data revealed a fault code that indicated that the onboard battery was permanently disabled due to a pack over-voltage condition, causing its safety monitor firmware to permanently disable its output.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4).
|
|
Event Description
|
The customer reported that the patient reported that the freedom onboard battery rapidly loses its charge.The customer also reported that the patient's onboard battery was exchanged.There was no reported patient impact.
|
|