CLEARSTREAM TECHNOLOGIES LTD. SLEEK RX; PTA BALLOON DILATATION CATHETER Back to Search Results
Catalog Number 425-4015X
Device Problems Component Falling (1105); Torn Material (3024)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/01/2016
Event Type  Injury  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.Receipt of the device is pending.The investigation is currently in progress.Not returned.
 
Event Description
It was reported that the device separated.The doctor was ballooning the popliteal artery with csi-jetstream and complaint device balloon tore off the balloon catheter.This separated piece had to be retrieved using a snare.The patient had no complications as a result of this failure and the procedure was completed.There was no patient injury reported.
 
Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.It was reported that the device separated.The device was flushed and prepped correctly prior to use.The operation was a straight pta of the distal leg.The target lesion was the popliteal artery.The lesion was highly calcified.The doctor was using a csi-jetstream.The device was delivered to the lesion with difficulty and inflated several times.The device was attempted to be removed from the patient however resistance was felt and the balloon separated from the shaft.This separated piece had to be retrieved using a snare.The device was inflated twice and deflated without issue.The patient had no complications as a result of this failure and the procedure was completed.There was no patient injury reported.The lot history records have been reviewed with special attention to the manufacturing and inspection of this product.The device was found to have met specifications prior to shipment.No manufacturing anomalies were identified that may have caused or contributed to the reported event.This is the first event reported for this lot number and issue to date.The device was not returned for evaluation.The result of the investigation is inconclusive as the complaint sample was not returned for evaluation.Based upon the available information a definitive root cause has not been determined.It is unknown if patient factors, handling or procedural techniques may have contributed to the reported event.Based on analysis performed no additional action is required at this time.Further to this the confirmed complaint rate for this failure mode is 0.015%.This is above the predicted rate of 0.015%.An event was already raised in our quality system to address this.The ifu states: description: ¿ the sleek® percutaneous transluminal angioplasty (pta) peripheral catheter family comprise a range of sizes of rapid exchange catheters for peripheral angioplasty.The catheter¿s proximal tubing is 304v stainless steel and the distal coaxial tubings are nylon co-polymer blend.The lumen of the shaft is used for the purpose of inflating and deflating the balloon.A second lumen at the tip is used for advancing the guidewire.Indications: the sleek® catheters are intended for balloon dilatation of the femoral, popliteal and infra-popliteal arteries.These catheters are not designed to be used in the coronary arteries.Warnings: ¿ reuse, resterilization, reprocessing and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Precautions: ¿ carefully inspect the catheter prior to use to verify that the catheter has not been damaged during shipment and that its size, shape and condition are suitable for the procedure for which it is to be used.Do not use if product damage is evident.¿ proper functioning of the catheter depends on its integrity.Care should be used when handling the catheter.Damage may result from kinking, stretching, or forceful wiping of the catheter.¿ if the hypotube kinks prior to or during use the catheter should be discarded.No attempt should be made to straighten a kink in the hypotube.Storage: store in a cool, dark, dry place.Use the catheter prior to the ¿use by¿ date specified on the package.Directions for use: inspection and preparation ¿ remove the balloon sleeve by first withdrawing the shipping mandrel slightly and then slowly removing the sleeve while holding the catheter as close to the balloon as possible.¿ if any resistance is felt, or if any stretching of the catheter is observed while removing the balloon sleeve, the product should not be used.¿ the catheter should then be inspected for bends, kinks or stretched portions.Do not use if product damage is evident.Deflation and withdrawal ¿ simultaneously withdraw the dilatation catheter and guidewire from the guiding catheter/sheath.As the balloon exits the vessel, use a smooth, gentle, steady, motion.If resistance is felt upon removal then the balloon, guidewire and the guiding catheter/sheath should be removed together as a unit under fluoroscopic guidance, particularly if balloon rupture or leakage is known or suspected.This may be accomplished by firmly grasping the balloon catheter and guiding catheter/sheath as a unit and withdrawing both together, using a gentle twisting motion combined with traction.(b)(4).
 
Event Description
It was reported that the device separated.The device was flushed and prepped correctly prior to use.The operation was a straight pta of the distal leg.The target lesion was the popliteal artery.The lesion was highly calcified.The doctor was using a csi-jetstream.The device was delivered to the lesion with difficulty and inflated several times.The device was attempted to be removed from the patient however resistance was felt and the balloon separated from the shaft.This separated piece had to be retrieved using a snare.The device was inflated twice and deflated without issue.The patient had no complications as a result of this failure and the procedure was completed.There was no patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLEEK RX
Type of Device
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford
EI 
Manufacturer (Section G)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford
EI  
Manufacturer Contact
fiona nimhullain
539237111
MDR Report Key5889909
MDR Text Key52557857
Report Number9616666-2016-00019
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
PMA/PMN Number
K072947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 08/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2018
Device Catalogue Number425-4015X
Device Lot Number50127506
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/07/2016
Initial Date FDA Received08/19/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/31/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/18/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-