ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) FORTIFY VR, U1.6 DF1 ID; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD1233-40
Device Problems Premature Discharge of Battery (1057); Difficult to Interrogate (1331)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/01/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the patient received a shock from an external source and presented at the hospital.The device was unable to be interrogated at the hospital.The device was explanted and replaced.The patient was stable post procedure.
 
Manufacturer Narrative
No conclusion code available.Final analysis found the reported inability to communicate was confirmed in the laboratory and was due to premature battery depletion.Based on all available parameter and usage information, device longevity was found to be below the expected limits.With an external power supply attached, the device functioned normally.No high current was detected during testing and the power consumption was normal.The original battery was returned to the vendor for further evaluation and analysis could not conclusively determine a root cause for the premature battery depletion.
 
Manufacturer Narrative
Premature battery depletion was confirmed by analysis.Bench testing on the device was performed, and no sources of high current were noted.In further analysis of the battery, lithium clusters were observed, but at the time of analysis the clusters did not appear to be in a location or size that would be sufficient to cause an internal short of the battery.As a result, the cause of the premature battery depletion could not be conclusively determined.However, from these analyses, in the absence of other root causes, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY VR, U1.6 DF1 ID
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jaime chavez
645 almanor avenue
sunnyvale, CA 94085
8184934022
MDR Report Key5893149
MDR Text Key52630567
Report Number2938836-2016-10355
Device Sequence Number1
Product Code LWS
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date 06/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2015
Device Model NumberCD1233-40
Device Lot Number4251055
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/01/2016
Initial Date FDA Received08/22/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
06/30/2017
Supplement Dates FDA Received09/27/2016
06/02/2017
06/30/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/16/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0115-2017
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-