• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG ROTAFLOW CENTRIFUGAL PUMP SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG ROTAFLOW CENTRIFUGAL PUMP SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number ROTAFLOW DRIVE
Device Problems Device Displays Incorrect Message (2591); No Flow (2991); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/31/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device will be evaluated by a maquet field service technician.A follow-up medwatch will be submitted if additional information becomes available.
 
Event Description
It was reported that the rotaflow got the signal error.The unit was reset and the flow stopped and the device started making a noise.Additional information: the incident occurred during patient treatment, no negative consequences for the patient or a delay in therapy were reported.(b)(4).
 
Manufacturer Narrative
A maquet field service technician was on site and investigated the unit.The technician determined that the root cause is the drive unit.The console operated as it should.A supplemental medwatch will be submitted if additional informations becomes available.
 
Event Description
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAFLOW CENTRIFUGAL PUMP SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key5893605
MDR Text Key53600584
Report Number8010762-2016-00534
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K991864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROTAFLOW DRIVE
Device Catalogue Number701022161
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/03/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2009
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-