It was reported that the battery reciprocator device did not power up with a full battery.During an in-house engineering evaluation, it was observed that the device was not working and had no power with a sticky trigger, frozen gear, corroded needle bearing and worn trigger components.It was not reported if the device was used in surgery, or if there was patient involvement reported.It was not reported if there were any delays in a scheduled surgical procedure or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|