Device evaluated by mfr: returned product consisted of the watchman access sheath (was).Blood was on the device and the valve was opened as received.The hub, shaft, and tip were examined.Microscope examination of the valve did not reveal any damage or irregularities; however, it was noted that the threads of the was were damaged/cross threaded.It could not be determined when the thread damage occurred.Multiple kinks were found on the shaft of the device.The tip was damaged with inner liner delamination.Functional testing of the valve confirmed the ability to completely close the valve with minimal forward force.Water was injected into the was by attaching a syringe filled with water.There were no leaks or air bubbles observed during water injection.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
|