(b)(4).A conclusion code could not be chosen.The complaint was confirmed, but the root cause is unknown.Upon receipt of the inspiratory limb of the infant breathing circuit it was visually examined for any signs of abuse/misuse/damage.The complaint has been confirmed.The temperature probe connector has broken away from the circuit inline "t" union.A device history record review could not be conducted since the lot number was not provided.The complaint has been confirmed.Root cause cannot be established.The breakage could be the result of too much force being applied either during removal or installation of the circuit temperature probe.Trending does not indicate the issue is wide spread.No further action required.
|