TELEFLEX MEDICAL HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE Back to Search Results
Catalog Number 031-33J
Device Problem Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 08/15/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A visual, functional, and dimensional inspection of the product involved in the complaint could not be conducted since the product was not returned.The device history record of the product 031-33j batch number 74d1601769 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No nonconformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled and inspected according to our specifications.No conclusion can be established at this time based on the lack of the device sample.Customer complaint cannot be confirmed due to the lack of the device sample to perform a proper investigation and determine the root cause.If the device sample becomes available this investigation will be updated with the evaluation results.
 
Event Description
The customer alleges that the adaptor couldn't connect with the oxygen flow meter.A new unit was used.
 
Manufacturer Narrative
(b)(4).The sample was returned for evaluation.A visual exam was performed and it was observed that the internal threads of the adaptor were damaged.Based on the investigation performed, the reported complaint was confirmed.The root cause for the issue was found to be the positioning of the thread lead and the softness of the new resin used for the snap adaptor.A capa was opened to address this issue.
 
Event Description
The customer alleges that the adaptor couldn't connect with the oxygen flow meter.A new unit was used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE
Type of Device
NEBULIZER ADAPTOR
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key5912061
MDR Text Key53415729
Report Number3004365956-2016-00329
Device Sequence Number1
Product Code CAF
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/29/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date04/21/2021
Device Catalogue Number031-33J
Device Lot Number74D1601769
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received09/26/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/15/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-