A customer reported to biomerieux that a fire started within the uninterruptible power supply (ups) attached to the computer for the vitek 2 instrument.The customer initially stated that the ups caught on fire.It was extinguished by laboratory personnel using a fire extinguisher.Upon further interview with the customer, it was noted that no flames were seen coming from the ups; however, smoke was noticed.Therefore the laboratory personnel used the fire extinguisher to extinguish it.No injury to laboratory personnel was incurred as a result of this event.No damage to the vitek 2 instrument nor the computer was noted as a result of the fire.Therefore, there was no impact to patient results nor a delay in results.An investigation will be initiated by biomerieux to investigate this event.
|
A customer reported to biomerieux that a fire started within the uninterruptible power supply (ups) attached to the computer for the vitek® 2 instrument, an internal biomereux investigation was performed.Results are as follows: the investigation was initiated in response to a customer complaint reporting that their ups 1000 emitted smoke.After review of the returned ups, it was noted that a capacitor failure was the cause of the smoke.It is not unusual for a failing capacitor to emit smoke prior to it's self-destruction and removal from the power source.There was no actual danger to the operator.Complaint history and trend review do not suggest a systemic quality issue.
|