The companion 2 driver was returned to syncardia for evaluation.Visual inspection of the external components revealed no abnormalities.Review of the electronic patient data file revealed no anomalies.During visual inspection of the internal components, the left compressor was removed and inspected.Visual inspection revealed that the counterweight was making contact with the inner surface of the end cap.The left compressor was reinstalled and when the companion 2 driver was tested, the reported grating noise was confirmed.The root cause of the grating noise was mechanical interference between the counterweight and the end cap of the left compressor.The left compressor was replaced, and the issue was resolved.No grating noise was heard, and the companion 2 driver passed all final performance testing.This failure mode posed a low risk to the patient because it did not prevent the companion 2 driver from performing its life-sustaining functions.The companion 2 driver is equipped with two compressors, a primary and a redundant, secondary compressor.In the event of a primary compressor malfunction, the companion 2 driver will switch to operation on the secondary compressor.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4).
|