Device evaluated by mfr: returned product consisted of the watchman access sheath (was).Blood was on the device and the valve was opened as received.Kinks were found 5cm, 26cm, and 52cm from the tip of the access system.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.Functional testing of the valve confirmed the ability to completely close the valve with minimal forward force.Water was injected into the was by attaching a syringe filled with water.There were no leaks or air bubbles observed during water injection.The reported issue was not confirmed.There was no evidence of any material or manufacturing deficiencies contributing to the reported event.Because there was no evidence of any product quality deficiencies, it was considered likely that the shaft kinks were attributable to procedural factors.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
|