• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS EMAX 2 PLUS MOTOR; MOTOR, DRILL, ELECTRIC - HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS EMAX 2 PLUS MOTOR; MOTOR, DRILL, ELECTRIC - HANDPIECE Back to Search Results
Catalog Number EMAX2PLUS
Device Problems Crack (1135); Overheating of Device (1437); Component Missing (2306); Defective Device (2588); Protective Measures Problem (3015)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Additional narrative: (b)(4) (b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported by (b)(6) that during service and evaluation, it was discovered that there was damaged component to the cable/cord/wiring of the motor device.It was further determined that the control had a crack, the motor (driver) and the coupling were defective.It was observed that the hose was worn; the device made metallic noises, the connector cap cover wire cable was missing and the temperature was 119 f to "3 unit s".It was further determined during the pre-repair diagnostics assessment that the device failed for loctite & cable document assessments, handpiece temperature assessment, hand control assessment, noise assessment and for safety assessment.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMAX 2 PLUS MOTOR
Type of Device
MOTOR, DRILL, ELECTRIC - HANDPIECE
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key5922965
MDR Text Key53797586
Report Number1045834-2016-12611
Device Sequence Number1
Product Code HBC
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K080802
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 07/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEMAX2PLUS
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/05/2016
Initial Date FDA Received09/01/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/25/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-