• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES MONUMENT DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS; GAUGE, DEPTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES MONUMENT DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS; GAUGE, DEPTH Back to Search Results
Catalog Number 319.006
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
No reported patient or surgical involvement.Device is an instrument and is not implanted or explanted.Subject device has been received; no conclusions could be drawn as the device is entering the complaint system.Service history review: no service history review can be performed as part 319.006 with lot 7971080 is a lot/batch controlled item.The manufacture date of this item is april 8, 2015, which is the release to warehouse date.The service history review is unconfirmed.Device history record review: manufacturing location: (b)(4) - manufacturing date: april 8, 2015.No non-conformance reports were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Device used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that a metal depth probe component has broken off of a depth gauge instrument.The issue was discovered during sterile processing as the device was dismantled for cleaning.There was no reported patient or procedural involvement.This report is 1 of 1 for (b)(4).
 
Manufacturer Narrative
An investigation summary was performed: the investigation of the complained this particular depth gauge is part of at least 14 technique guides, including the 2.4 mm variable angle lcp distal radius system and is used to measure the depth of the holes for the 2.0mm/2.4mm screws to ensure the correct screw length is used during the procedure.The information is provided per the 2.4 mm variable angle lcp distal radius system technique guide.Unable to determine a definitive root cause.However, the complaint condition was most likely caused by rough handling during sterilization.It is not likely that the design of the device contributed to this complaint.Service and repair evaluation: the customer reported the metal tip broke off the depth gauge.The repair technician reported the tip broke off, and the slide sleeve was missing.Tip broken is the reason for repair.The item is not repairable per the inspection sheet.The cause of the issue is unknown.The item will be forwarded to customer quality.The evaluation was confirmed.Device used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS
Type of Device
GAUGE, DEPTH
Manufacturer (Section D)
SYNTHES MONUMENT
1051 synthes ave
monument CO 80132
Manufacturer (Section G)
SYNTHES MONUMENT
1051 synthes ave
monument CO 80132
Manufacturer Contact
mark vornheder
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5925328
MDR Text Key53906897
Report Number1719045-2016-10663
Device Sequence Number1
Product Code HTJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number319.006
Device Lot Number7971080
Other Device ID Number(01)10886982189943(10)7971080
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/19/2016
Initial Date FDA Received09/02/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/12/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/08/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-