Catalog Number 160176
Device Problems Break (1069); Fracture (1260); Material Integrity Problem (2978)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 08/08/2016
Event Type  malfunction  
Manufacturer Narrative
When completed, the investigation results will be submitted in a supplemental report.
 
Event Description
Tibial insert impactor broke during surgery.Doctor (b)(6) was hitting with a mallet and it snapped.
 
Manufacturer Narrative
An event regarding crack/fracture involving an mako impactor was reported.The event was confirmed.Method & results: device evaluation and results: visual inspection shows a sheared break and some portion of broken insert was stucked in the upper side of impactor.Review of device by material analysis engineer indicated: "the damage on the impaction pad is consistent with in-service use." medical records received and evaluation: no medical records or x-rays were made available for evaluation.Device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.Complaint history review: there has been no other event for the lot referenced.Conclusions: visual inspection shows a sheared break and some portion of broken insert was stucked in the upper side of impactor.Review of device by material analysis engineer indicated that the damage on the impaction pad is consistent with in-service use.No further investigation for this event is possible at this time.If additional information will be received, this investigation will be reopened and re-evaluated.
 
Event Description
Tibial insert impactor broke during surgery.Doctor (b)(6) was hitting with a mallet and it snapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL INSERT IMPACTOR
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
cindy chuhinko
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5925853
MDR Text Key54524358
Report Number3005985723-2016-00277
Device Sequence Number1
Product Code NPJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/02/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number160176
Device Lot Number12240410
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received01/25/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/21/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age71 YR
-
-