Catalog Number CSR60
Device Problems Bent (1059); Vibration (1674); Scratched Material (3020)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to wear/strained from normal use and servicing.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported by (b)(6) that during service and evaluation, it was discovered that the drive shaft was bent on the compact speed reducer device.It was noted on the service order that the drive shaft was scratched and there was vibration.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORATOR DRIVER W/HUDSON END
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES - COMPACT SPEED REDUCER
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key5928723
MDR Text Key54049259
Report Number1045834-2016-12626
Device Sequence Number1
Product Code HBE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K042783
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 08/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/06/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCSR60
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received08/23/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/11/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-